Deep clean

The Deep clean runs for 12 hours.

Step-by-step

1

Switch to the deep clean vial

Transfer tubing kit to the deep clean vial.

2

Start the Deep clean program

Press and hold the Measure button for 2 seconds to enter the Programs menu. Use the navigation buttons (OK ) to select the Deep clean program. Press OK to start the Deep clean program.

3

Start a clean Clean program

When the Deep clean program has completed, transfer tubing kit to a fresh disinfection vial and run a Clean program. See Clean the setup.

4

That's it!

The Deep clean is complete.Discard the deep clean vial. It is for single use only.

Qualify your BactoBox® setup after the Deep clean program is done