Installation checklist

🚧 Work in progress 🚧

This page may look complete but work is still underway.

Parameter
Value

Company name

Department name

BactoBox® device ID

Item
Compliant
Not complaint
Date
Signature

There are no strong vibration sources near the BactoBox® setup.

There are no steam sources near the BactoBox® setup.

The humidity in the installation environment is always below 60 %

The temperature in the installation environment is always between +15℃ and +35℃

The air in the installation environment is always free from large dust particles. E.g., sawdust and the like.

There is a flow cell in the BactoBox® device.

The flow cell is firmly in place. That is, there is be a 2–3 mm gap between the clamps and the flow cell cartridge itself.

The flow cell cover is in place on top of the flow cell compartment

There is a tubing kit connected to the BactoBox® device. Both inlet (white) and outlet (black).

There is an external filter connected between the tubing kit and the BactoBox® device.

The external filter is connected to the inlet. That is, the white luer lock.

The power supply fits the BactoBox® device. That is, the power supply label says that its output is 12 V⎓ / 2500 mA / 30 W.

The power supply is connected to the BactoBox® device.

The power supply is connected to the wall socket.